Evaluating the Quality of Optimal MRCP Image Using RT-2D-Compressed SENSE(CS)Turbo Spin Echo: Comparing Respiratory Triggering(RT)-2D-SENSE Turbo Spin Echo and Breath Hold-2D-Single-Shot Turbo Spin Echo

Item request has been placed! ×
Item request cannot be made. ×
loading   Processing Request
  • Additional Information
    • Publication Information:
      MDPI AG, 2022.
    • Publication Date:
      2022
    • Collection:
      LCC:Computer applications to medicine. Medical informatics
    • Abstract:
      This study aimed to select the pulse sequence providing the optimal MRCP image quality by applying various reduction and denoising level parameters—which could improve image quality and shorten examination time—to BH-2D-SSh TSE, RT- 2D-SENSE TSE, and RT-2D-Compressed SENSE(CS) TSE and then comparing and analyzing the obtained images. This study was carried out using 30 subjects (15 men and 15 women with a mean age of 53 ± 8.76 years) who underwent an MRCP test using 3.0T MRI equipment. These 30 subjects were composed of 20 patients (CHDD: 7; LC: 6; and IPMN: 7) and 10 volunteers without a disease. When the CS technique was used, five reduction values (1.1, 1.2, 1.3, 1.4, and 1.5) were used and four denoising levels (No, Weak, Medium, and Strong) were used. The existing SENSE method was based on a reduction value of 1, and other parameters were set the same. The image data of BH-2D-SSh TSE, RT-2D-SENSE TSE, and RT-CS-2D TSE used for the analysis were acquired in the coronal plane, and the acquired data underwent MIP post-processing for analysis. To compare these techniques, SNR and CNR were measured for six biliary duct images for the purpose of quantitative analysis, and qualitative analysis was performed on the sharpness of the duct, the overall quality of the image, and the motion artifact. The results of the quantitative and standard analyses showed that the RT-2D-CS TSE technique had the highest results for all IPMN, LC, and CHDD diseases (p < 0.05). Moreover, SNR and CNR were the highest when the reduction value was set to 1.3 and the denoising level was set to medium as the CS setting values (p < 0.05). Compared with the conventional RT-2D-SENSE TSE, the test time decreased by 20% and SNR and CNR increased by 14% on average. When conducting RT-2D-CS TSE, we found that it shortened the examination time and improved the image quality compared to the existing RT-2D-SENSE TSE. Unlike previous studies, this study using the RT technique shows that it is a useful MRI Pulse Sequence technique able to replace the BH-2D-SSh TSE and BH-3D-SENSE GRASE techniques, which require the patient to hold their breath during the test.
    • File Description:
      electronic resource
    • ISSN:
      2379-139X
      2379-1381
    • Relation:
      https://www.mdpi.com/2379-139X/8/3/111; https://doaj.org/toc/2379-1381; https://doaj.org/toc/2379-139X
    • Accession Number:
      10.3390/tomography8030111
    • Accession Number:
      edsdoj.2c83a6a6f12a421b9a0513eff3999862