Item request has been placed! ×
Item request cannot be made. ×
loading  Processing Request

Efficacy of Dural Sealant System for Preventing Brain Shift and Improving Accuracy in Deep Brain Stimulation Surgery.

Item request has been placed! ×
Item request cannot be made. ×
loading   Processing Request
  • Additional Information
    • Source:
      Publisher: Japan Neurosurgical Society Country of Publication: Japan NLM ID: 0400775 Publication Model: Print-Electronic Cited Medium: Internet ISSN: 1349-8029 (Electronic) Linking ISSN: 04708105 NLM ISO Abbreviation: Neurol Med Chir (Tokyo) Subsets: MEDLINE
    • Publication Information:
      Original Publication: Tokyo : Japan Neurosurgical Society.
    • Subject Terms:
    • Abstract:
      The success of deep brain stimulation (DBS) depends heavily on surgical accuracy, and brain shift is recognized as a significant factor influencing accuracy. We investigated the factors associated with surgical accuracy and showed the effectiveness of a dural sealant system for preventing brain shift in 32 consecutive cases receiving DBS. Thirty-two patients receiving DBS between March 2014 and May 2015 were included in this study. We employed conventional burr hole techniques for the first 18 cases (Group I) and a dural sealant system (DuraSeal) for the subsequent 14 cases (Group II). We measured gaps between the actual positions of electrodes and the predetermined target positions. We then retrospectively evaluated the factors involved in surgical accuracy. The average gap between an electrode's actual and target positions was 1.55 ± 0.83 mm in all cases. Postoperative subdural air volume e, the only factor associated with surgical accuracy (r = 0.536, P < 0.0001), was significantly smaller in Group II (Group I: 43.9 ± 27.7, Group II: 12.1 ± 12.5 ml, P = 0.0006). The average electrode position gap was also significantly smaller in Group II (Group I: 1.77 ± 0.91, Group II: 1.27 ± 0.59 mm, P = 0.035). Use of a dural sealant system could significantly reduce intracranial air volume, which should improve surgical accuracy.
    • References:
      Stereotact Funct Neurosurg. 2008;86(3):173-8. (PMID: 18334860)
      Neuromodulation. 2008 Oct;11(4):302-10. (PMID: 22151145)
      Spine (Phila Pa 1976). 2015 Apr 15;40(8):505-13. (PMID: 25646746)
      World Neurosurg. 2012 Nov;78(5):498-504. (PMID: 22381303)
      Spine (Phila Pa 1976). 2011 Nov 1;36(23):1906-12. (PMID: 22008746)
      J Neurosurg. 2007 Nov;107(5):983-8. (PMID: 17977271)
      Neurosurgery. 2010 Jul;67(1):49-53; discussion 53-4. (PMID: 20559091)
      J Neurosurg. 2007 Jan;106(1):52-8. (PMID: 17236487)
      Neurol Med Chir (Tokyo). 2013;53(1):1-6. (PMID: 23358161)
      Neurosurgery. 2005 Jul;57(1 Suppl):146-51; discussion 146-51. (PMID: 15987581)
      Stereotact Funct Neurosurg. 2008;86(1):44-53. (PMID: 17881888)
      J Neurol Neurosurg Psychiatry. 2003 Dec;74(12):1631-7. (PMID: 14638880)
      Neurosurgery. 2010 Sep;67(3 Suppl Operative):ons213-21; discussion ons221. (PMID: 20679927)
      Stereotact Funct Neurosurg. 2002;78(3-4):132-45. (PMID: 12652038)
      J Neurosurg. 2011 Nov;115(5):971-84. (PMID: 21800960)
      J Neurosurg. 2007 Nov;107(5):989-97. (PMID: 17977272)
      Neuromodulation. 2015 Oct;18(7):574-8; discussion 578-9. (PMID: 26222380)
      Stereotact Funct Neurosurg. 2008;86(1):37-43. (PMID: 17881887)
      Cent Eur Neurosurg. 2011 Nov;72 (4):181-5. (PMID: 21739410)
    • Contributed Indexing:
      Keywords: brain shift; deep brain stimulation; dural sealant system; surgical accuracy
    • Accession Number:
      0 (Drug Combinations)
      0 (DuraSeal spinal sealant)
      0 (Oligopeptides)
      3WJQ0SDW1A (Polyethylene Glycols)
    • Publication Date:
      Date Created: 20180501 Date Completed: 20181031 Latest Revision: 20181202
    • Publication Date:
      20250114
    • Accession Number:
      PMC5958041
    • Accession Number:
      10.2176/nmc.oa.2017-0242
    • Accession Number:
      29710057