Abstract: In this bachelor's thesis results of the optimization of the measurement of 4-nitrophenol (1·10-4 mol·l-1 ) in the Britton-Robinson (BR) buffer medium on the silver rotating disk electrode are presented with the aim of suppressing the passivation of the electrode surface. The parameters of the differential pulse voltammetry (DPV) technique were optimized for the analytical use of the rotating disk electrode. In BR buffer pH 4,0 and at the electrode rotation speed of 250 rpm calibration dependences of 4-nitrophenol was obtained in the range of (1-10) ·10-5 mol·l-1 under selected optimal conditions and different methods of electrode surface pretreatment. The limit of quantification (LOQ) achieved with polishing before each measurement was 1,7·10-6 mol·l-1 and with polishing only before the concentration change was 8,8·10-6 mol·l-1 . ; V této bakalářské práci jsou předloženy výsledky optimalizace měření 4-nitrofenolu (1·10-4 mol·l-1 ) v prostředí Brittonova-Robinsonova (BR) pufru na stříbrné rotující diskové elektrodě s cílem potlačit pasivaci povrchu elektrody. Pro analytické využití rotující diskové eleketrody byly optimalizovány parametry techniky diferenční pulzní voltametrie (DPV). V prostředí BR pufru pH 4,0 a při rychlosti rotace elektrody 250 rpm byly získány kalibrační závislosti 4-nitrofenolu v rozmezí (1-10) ·10-5 mol·l-1 . Dosažená mez stanovitelnosti (LOQ) s leštěním před každým měřením byla 1,7·10-6 mol·l-1 a s leštěním pouze před změnou koncentrace byla 8,8·10-6 mol·l-1 . ; Katedra analytické chemie ; Department of Analytical Chemistry ; Faculty of Science ; Přírodovědecká fakulta